| Quantum Efficiency at 20° C |
New VIS-NIR Optimized CCD (Also available: 532 nm optimized spectrometer) |
| CCD Sensor Format |
2048 x 70 |
| CCD Pixel Size |
14 x 14 µm |
| CCD Height |
1 mm |
| CCD QE at 20° C |
60% at 500 nm; 80% at 600 nm; 80% at 800 nm 68% at 900 nm; 42% at 1000 nm; 20% at 1,060 nm See curve below |
| Deep Thermoelectric Cooling |
-50° C @ +25° C ambient or -40° C @ +40° C ambient. Yields low dark current suitable for most OEM and many research applications |
| Single Pixel Well Capacity |
50,000 e¯/pixel (minimum); 60 000 e¯/pixel (typical) |
| Serial Register Full Well Capacity |
250,000 e¯/pixel (minimum); 500 000 e¯/pixel (typical output register saturation) |
| Scan Rates |
45 kHz and 500 kHz |
| Readout Noise (at 45 kHz and at –45° C) |
10 e¯ (typical) to 12 e¯ (maximum) |
| Readout Noise (at 500 kHz and at –45° C) |
20 e¯ (typical) to 25 e¯ (maximum) |
| Dynamic Range (typical) |
55,500:1 |
| Maximum Spectral Rate |
20 Hz at 45 kHz scan rate; 189 Hz at 500 kHz scan rate |
| Digitization |
16-bit ADC |
| Non-linearity (measured on each camera) |
<0.15% (typical) at 45 kHz (0.4% maximum) <0.20% (typical) at 500 kHz (1% maximum) |
Dark Current at –45° C Note: Pixel Size = 14 μm |
0.025 e–/pixel/s (typical) |
| Software-adjustable Gains |
2, 4, and 10 e–/count at –50° C |
| Communication |
USB 2 |
| Environmental Conditions |
Operating temperature 0° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
Power Requirements AC/DC Power Supply (provided) |
90-264 VAC, 47–63 Hz |