분광기, 분광 시스템, 그레이팅, 회절격자, 카메라 시스템, OEM 제조업체
UV-VIS-NIR Spectrometer
 
 
Horiba’s OEM division, a major supplier of OES solutions, has developed a high end OES Spectrometer with a proprietary optical configuration for broad coverage (185-1050 nm) and high resolution. The OES-Star is available with a selection of three UV-VIS-NIR Back-Illuminated (B.I.) sensors, one TE-Cooled and two uncooled CCD detectors. This unique OES system features an optical design that provides a high peak symmetry across a very broad spectral range combined with high spectral resolution better than 1 nm, at an unmatched aperture of F/2.3.
For OEM Industrial Applications
ㆍ Optical Emission Spectrometry
ㆍ Reflectometry
ㆍ Process and Quality Control
ㆍ Examples :
     - Endpoint Detection
     - Plasma Monitoring
     - Semiconductor Wafer Inspection
ㆍ Optical emission spectrometry (OES)
ㆍ Unmatched long term stability and excellent wavelength accuracy
ㆍ USB2 Ethernet and EtherCAT interface
ㆍ High spectral resolution combined with high throughput
ㆍ Broad wavelength coverage from UV to NIR 
ㆍ Multiple B.I. CCD options cooled/uncooled with high dynamic range 
ㆍ CMOS sensor option with electronic shutter and ultra high spectral rate
Flat resolution and dispersion across the entire spectral range (185 -1050 nm) combined with excellent peak symmetry

This OES-Star is featuring a tall CCD with built in round-to-line (RTL) fiber converter to maximize light collection and therefore the system throughput. Input fiber with a core of 1 mm or smaller can be used. However to take advantage of the tall detector and RTL, larger core fibers are recommended.
Built-in round-to-line (RTL) converter boosts throughput when combined with a large core diameter fiber and our tall CCD (S13241-1108S)

Quantum Efficiency Plot

OEM Spectrometer Selection Guide

| General Spectrometer Specifications* | |
|---|---|
| 
 Spectral Coverage  | 
185 – 1050 nm (2048 horizontal pixel format with 14 µm pixel size) 200 – 1050 nm (2048 horizontal pixel format with 12 µm pixel size)  | 
| 
 Spectrometer Input  | 
SMA (standard), other input types available upon request | 
| 
 Spectral Resolution  | 
<1 nm with 22 µm slit with very high symmetry across broad spectral range (flexible based on customer’s need)  | 
| 
 Average Spectral Dispersion  | 
34.6 nm/mm, 0.484 nm/pixel (14 µm pixel size), 0.41 nm/pixel (12 µm pixel size) | 
| 
 Focal Length  | 
116 mm | 
| 
 F/#  | 
F/2.3 | 
| 
 Wavelength Accuracy  | 
< +/- 0.1 nm | 
| 
 Stray Light  | 
< 0.1%* | 
| 
 Software  | 
 LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration)  | 
| CCD Detector Model | Cooled BI High-end CCD with best dynamic range (S13241-1108S)  | 
BI CCD Middle Grade (S14660-2048N)  | 
BI CCD cost effective (S11071-1106N)  | 
|---|---|---|---|
| CCD Sensor Format | 2048 x 256 | 2048 x 192 | 2048 x 70 | 
| CCD Pixel Size | 12 x 12 μm | 14 x 14 μm | 14 x 14 μm | 
| Image Size | 24.5 x 3.0 mm | 28.6 x 2.68 mm | 28.7 x 1.0 mm | 
| CCD QE at 20°C | 90% or higher at the peak | High sensitivity over a wide spectral range  | 
> 75% for 450 - 750 nm | 
| Deep Thermoelectric Cooling | -5º C | Non-cooled | Non-cooled | 
| Full Well Capacity | 500,000 e¯/pixel (typical) | 200,000 e¯/pixel (typical) | >200,000 e¯/pixel (typical) >175,000 e¯/pixel (minimum)  | 
| Readout Noise | 50 e¯ (typical) (@ 1.5MHz sampling)  | 
40 e¯ (typical) 50 e¯ (maximum)  | 
35 e¯ (typical) 45 e¯ (maximum)  | 
| Digitization | 16-bit ADC | 16-bit ADC | 16-bit ADC | 
| Dynamic range (FW/RN) | 10000:1 | 5000:1 | 5700:1 | 
| Non-linearity (measured on each system)  | 
<1.8% (maximum) at low gain <0.9% (maximum) at high gain  | 
<0.1% (corrected) | <0.1% (corrected) | 
| Dark Current | 1.5 e-/pixel/s (typical) @-5° C 15 e¯/pixel/s (maximum)  | 
50 e¯/pixel/s (typical) @ 20° C 500 e¯/pixel/s (maximum)  | 
50 e¯/pixel/s (typical) @ 20° C 500 e¯/pixel/s (maximum)  | 
| Environmental Conditions | Operating temperature 0° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -20° C to 50° C  | 
||
| Power Requirements AC/DC Power Supply (provided)  | 
90-264 VAC, 47–63 Hz | ||
*Specifications, form factor, and spectrometer cover subject to change without notice. No LabVIEW™ license is needed to run our acquisition software.