High Spectral Rate and High Dynamic Range Miniature Spectrometers

VS70-HSR & VS70-HDR Optical Layout VS70-HSR & VS70-HDR Spectral Coverage VS70-HSR & VS70-HDR Quantum Efficiency

VS70 is a high performance compact fiber coupled spectrometer system covering wide spectral range of 200 to 1050 nm. This OEM optical module is built around Horiba’s type-IV Aberration-Corrected Flat-Field Holographic Ion-Etched concave grating and is specially designed to easily adapt to a large variety of detectors and electronic drivers. This system is fitted with a custom variable order-sorting filter to eliminate higher orders.

The proprietary layout of the VS70 features a single-optics design for superior imaging, peak symmetry and resolution, sensitivity and low-stray light performance, for low-light applications (fluorescence, emission) and high-dynamic applications (absorbance, reflectance). 

 

For OEM Industrial Applications

Ideal for industrial low-light applications such as fluorescence, emission, absorbance, and reflectance.

Examples:

ㆍSemiconductor plasma monitoring
ㆍColorimetry
ㆍBlood and urine analysis
ㆍLiquid chromatography
ㆍFood moisture content analysis
ㆍFlow cytometry

ㆍHigh spectral resolution and system throughput
ㆍCompact size robustness and stability
ㆍHigh Spectral Rate (HSR)
ㆍHigh Dynamic Range (HDR)
ㆍWide spectral covrage from UV to NIR combined with ultra low stray light
ㆍHigh linearity (raw and linearity-correction
ㆍHighly customizable for various applications

 

OEM Spectrometer Selection Guide

 

 

General Spectrometer Specifications*

Spectral Coverage

Standard: UV-VIS (200–860 nm) with 477 l/mm grating, VIS (380–750 nm) with 582 l/mm
grating, UV-NIR (200–1050 nm) with double-blazed 365 l/mm grating
On request other spectral ranges available; different mechanical integration might needed

Spectrometer Input

Fiber coupled: FC or SMA; Free space: 12-25-37-50-62-75-100-125-150-200 μm slits, other options
available upon request

Spectral Resolution

2 nm for 365 l/mm grating, 1 nm for 477 l/mm grating, 0.8 nm for 582 l/mm grating
(25 μm slit for all)
Average Spectral
Dispersion
30.9 nm/mm for 365 l/mm grating, 23.7 nm/mm for 477 l/mm grating
16.3 nm/mm for 582 l/mm grating

Focal Length

70 mm

Options Selection of high grade sensors: CMOS, B.I. CCD, PDA
Input port: SMA, FC, free space, custom input

F/#

F/2.8
Stray-light Rejection
Typical (Maximum)
0.01% (0.02%) for UV-VIS configuration with 300 μm tall CCD, measured at 700 nm
(measured with broad bandpass 510 nm filter, 75 μm slit-width)
> 2.4 AU linear range (5% variation) with caffeine 273 nm absorption peak in 10 mm cuvette
and D2 lamp

Wavelength Accuracy

0.2 nm

Software**

LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration)

 

 

Detector Options and Specifications*
Linear back-illuminated CCD with large active area and high sensitivity
Detector Model Hamamatsu CCD sensor S11071 with high
spectral acquisition speed
Hamamatsu CCD sensor S10420 with high
dynamic range
Sensor Format 2048 x 70 pixels, shorter version
available on request
2048 x 70 pixels, shorter version
available on request
CCD Pixel Size 14 x 14 μm 14 x 14 μm
CCD Active Area 28.7 x 1 mm 28.7 x 1 mm
CCD QE > 75% for 450-750 nm > 75% for 450-750 nm
Sensor Temperature Uncooled Uncooled
Maximum Spectral Rate 760 spectra/s 223 spectra/s
Full Well Capacity >200,000 e¯ (typical); >175,000 e¯ (minimum) >225,000 e¯ (sensitivity mode); >375,000 e¯
(high full well mode)
Readout Noise 35 e¯ (typical); 45 e¯ (maximum) 35 e¯ (typical) & 45 e¯ (maximum) in
sensitivity mode; 50 e¯ (typical) and
75 e¯ (maximum) in high full well mode
Digitization 16-bit 16-bit
Dynamic Range (FW/RN) 6,000:1 (typical) 7,000:1 (typical) in sensitivity mode;
5,500:1 (typical) in high FW mode
Non-linearity (measured
on each system)
< 0.4% (corrected) < 0.4% (corrected)
Dark current @ (25°C) 50 e-/pix/s (typical); 500 e-/pix/s (maximum) 7 e-/pix/s (typical); 70 e-/pix/s (maximum)
Communication USB2 USB2
Environmental Conditions Operating temperature 15° C to 40° C ambient
Relative humidity <70% (non-condensing)
Storage temperature -25° C to 45° C
Power Requirements Molex connector (24 VDC)

 

* HORIBA Instruments has a policy of continuous product development, and reserves the right to amend part numbers, descriptions and specifications without prior notice.

**No LabVIEW™ license is needed to run our acquisition software.

 

Quantum Efficiency

 

 

Mechanical Drawings