product Image
DiaPlexQ™ MTC/NTM Detection Kit (Tuberculosis)
Model Name
-
Series
Molecular Diagnostic (Pathogen Detection)
Data
-
Manufacturer information
provider logo
SOLGENT

DiaPlexQ™ Novel Coronavirus (2019-nCoV) Detection Kit, Covid19, Corona virus

Problem with product info?
Update request
Manufacturer
SOLGENT
Product Type
Machine
Brand
-
SKU
125208
Product Name
DiaPlexQ™ MTC/NTM Detection Kit (Tuberculosis)
Model Name
-
Size
-
Weight
-
Product Details

DiaPlexQ™ MTC/NTM Detection Kit

Real-Time PCR based assay system for simultaneous detection of MTC and NTM complex

CE-IVD


M y cobacterium tuberculosis (MTB) is a pathogenic bacteria species and the causative agent in most cases of tuberculosis. Along with MTB, 5 other species, which we call MTC are regarded as common infectious agents causing tuberculosis. That is why it is necessary to detect MTC, along with MTB. Although NTM is a group of non-tuberculosis related mycobacteria, it usually causes lung infections that mimic tuberculosis resulting in a potential misdiagnosis. Thus, the identification of a single infection or co-infection of MTC and NTM is required for appropriate treatment.




Specification


DiaPlexQ™ MTC/NTM Detection KitDiaPlexQ™ MTC/NTM Detection Kit
ver 3.0
DiaPlexQ™ MTC/NTM Detection Kit
ver 4.0
Detection targetMTC (2 species): IS6110
NTM (13 species): 16S rRNA
MTC (2 species): IS6110, MPB64
NTM (15 species): 16S rRNA
MTC (2 species): IS6110, MPB64
NTM (6 species): 16S rRNA
RegistrationKFDA, CE-IVDCE-IVDCE-IVD
Detection technologyReal-Time PCR
Specimen typeNasopharyngeal swab, Nasopharyngeal aspirate, Bronchoalveolar lavage (BAL)
Analytical sensitivity10 copies
Compatible instruments*ABI 7500 / 7500 Fast Real-Time PCR System (Applied Biosystems)
CFX96™ Real-Time PCR System (Bio-Rad)
PCR running time~ 1 hr 20 min




Features

- HotStart PCR system : Ultra high specific and sensitive result

- DnaFree™ system : No host genomic DNA contamination ( DiaPlexC ™)

- UDG system : No carryover contamination

- Multiplex PCR system : Multiple targets in a single reaction

- Reliable system : Automatic PCR control

- Positive control included

- DNA extraction solution included (Cat. No. SQD21-K100)

- Easy-to-use master mix

- CE certification





Experimental




Sample typeFAMHEXCy5Result
Positive Control+++Valid
Negative Control--+Valid
NTC(Non-Template Control)--+Valid
Sample case 1+++/-MTC, NTM (Co-infection)
Sample case 2+-+/-MTC
Sample case 3-++/-NTM
Sample case 4--+Negative
Sample case 5---Required   re-experiment



Citation&Papers

Korean J Clin microbial, Seong Deok Lee, Hye Young Lee, Hyun Chul Kim, Soo Young Kim,

Mycobacterium tuberculosis and Non-Tuberculous Mycobacteria by PCR Assay.

Ryan KJ, Ray CG (Editors) (2004).

Sherris Medical Microbiology (4th ed.). McGraw-Hill. Clinical Characteristics of Non-Tuberculous Mycobacterial Pulmonary Disease in a National Tuberculosis Hospital.

Sun-Pil Choi, M.D., Bong-Keun Lee M.D.1, Jin-Hong Min, M.D., Jin-Hee Kim, M.D.

Pathogenic Classification and Clinical Charateristics of Non-Tuberculous Mycobacterial Pulmonary Disease in a National Tuberculosis Hospital.





Ordering information

TechnologyCat. No.Product
Real-time PCRSQD21-K100DiaPlexQ™ MTC/NTM Detection Kit (w/Ext.)
SQD20-K100DiaPlexQ™ MTC/NTM Detection Kit
SQD26-K100DiaPlexQ™ MTC/NTM Detection Kit (w/Ext.)- Ver.3.0
SQD25-K100DiaPlexQ™ MTC/NTM Detection Kit- Ver.3.0
SQD28-K100DiaPlexQ™ MTC/NTM Detection Kit (w/Ext.)- Ver.4.0
SQD27-K100DiaPlexQ™ MTC/NTM Detection Kit- Ver.4.0



SOLGENT
SOLGENT
DiaPlexQ™ Novel Coronavirus (2019-nCoV) Detection Kit, Covid19, Corona virus
Inquiry