| Product name |
Cylinder pump |
| Model name |
Anyfusion V-100 |
| Flow rate |
0.1 ~ 999.9 ml/hr |
| Flow rate accuracy |
±2% |
| VTBI selector, ml |
0.1 ~ 9999ml |
| Total Volume |
0.01 ~ 9999ml |
| KVO rate |
0.1 ~ 5 ml/hr |
| IV set |
Compatible with any brand of IV sets
(Compatible with installed micro filter 0.2~5.0um)
|
| Purge rate, ml/hr |
700 ml/hr |
| Bolus rate, ml/hr |
500(default), 1 ~ 999.9 ml/hr |
| Bolus volume |
3(default), 0.11 ~ 99.9 ml/hr |
| Occlusion pressure |
150 ~ 750 mmHg |
| Display |
3.5" TFT-LCD (Resolution : 240 x 320) |
| Alarm |
|
- Infusion completed alarm
- Start reminder alarm
- Repeat alarm
- Low battery alarm
- No battery alarm
- Depleted battery alarm
|
- Occlusion alarm
- Air-in-Line alarm
- Door open alarm
- System malfunction alarm
- Cylinder detached alarm
-Input time expired alarm
|
|
| Power |
AC100 ~ 240V, 50/60Hz / Fuse : 250V, 1.6A |
| Batter type |
Rechargeable Li-ion |
| Battery life |
Approx, 8hrs at 5ml/h, 10hrs at 25ml/h, 7hrs at 999.9ml/h |
| Recharge time |
Approx. 5hrs |
| Power consumption |
Max. 65VA |
| Protection type |
- Protection from electric shock : Class 1
- Protection capacity for electric shock : Type CF
- Protection from liquid infiltration : IPX2
|
| Dimension |
102(W) x 147(D) x 204(H)mm |
| Weight |
Approx 1.9kg |
| Operating condition |
- Temperature : 10~40โ
- Humidity : 30~75%
- Atmospheric pressure : 700 ~ 1,060hpa
|
| Storage condition |
- Temperature : -20~60โ
- Humidity : 10~95%
- Atomosheric pressure : 650 ~ 1,060 hpa
|
| Accessories |
Syring holder, Pole clamp, AC power cord, Operator manual |