The H2S/TOS TS MEDOR® is an autoGC-ED (MEDOR® Electrochemical wet cell Detector) for the analysis and monitoring of H2S, TOS and TS.
Different configurations exist depending on the application and concentration range : HS2/TOS TS MEDOR ppb for 0-1 ppm and H2S/TOS TS MEDOR ppm for 0-10 / 0-50 / 0-100 ppm
Principle
Automatic sampling using a loop
Automatic loop injection on metallic capillary column
Isothermal gas chromatograph
Detection of all compounds eluting from the column performed by MEDOR® detector : Electrochemical wet cell Detector which is a Sulfur Specific Detector (SSD)
Signal provided by electrochemical reaction between the wet cell electrolyte and the sulfur compounds
Fully compliant with ASTM D 7493-14 : Standard Test Method for On-line Measurement of Sulfur Compounds in Natural Gas and Gaseous Fuelsby Gas Chromatograph and
Electrochemical Detection.
Compliant with ASTM D 5504-20 : Standard Test Method for Determination of Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and
Chemiluminescence with alternative detector.
Key points
Continuous monitoring with automatic online sampling
Analytical performances : - Specific, linear and very sensitive to sulfur compounds - Results validation by automatic standard injection - Long term stability using detector installed in reservoir
Extremely low maintenance : - Very long life time detector with electrolyte, up to 10 years. - Low gas consumption, can be reduced in option - More than 10 years data storage - No cylinders required thanks to internal calibration tube and gas generators
Automatic control with process device
Intelligents system with tunable and interactive alarm levels
Powerful VISTACHROM Chromatotec® software : - Remote monitoring & injection control - Full traceability with on board archiving of results and chromatograms