Multichannel Grating Spectrometer
 
 
Horiba’s OEM Polispectra Quad spectrometer is an integration of four VS70 spectrometers built around Horiba JY Type-IV flat-field aberration-corrected holographic concave gratings. This high throughput VIS spectrometer with 4 fiber inputs provides simultaneous acquisition of 4 channels at high speed with minimal crosstalk. The main electronics board is designed to read the two CCDs on the system simultaneously and provides 1-8X binning capability. This OEM system is a unique and innovative approach to drastically reduce the cost per channel.
For OEM Industrial Low-light Applications
ㆍ Fluorescence 
ㆍ Emission (OES)
ㆍ Absorbance
ㆍ Reflectance
ㆍ Examples :
     - Body Fluids Analysis 
     - Flow Cytometry
     - LED Testing, Colorimetry
     - Food Analysis Moisture Content
 
ㆍ Simultaneous acquisition of 4 channels
ㆍ Excellent light purity with a typical 0.05% channel to channel crosstalk
ㆍ Spectral acquisition time as low as 1.3 ms for all 4 channels 
ㆍ High throughput (f/2.3)
ㆍ Ultra-low stray light with aberration correction holographic grating
ㆍ High SNR and dynamic range
ㆍ TTL in/out to trigger each and all channels
ㆍ Compact size, robustness, stability and affordability 
CCD Quantum Efficiency

Simultaneous measurement of 4 colors LED array spectra with Polispectra Quad spectrometer

Polispectra Quad Spectrometer is capable of simultaneously measuring 4 channels, the picture above shows the emission spectra of 4 different color LEDs measured using Ø 200 µm core fibers
Raw Throughput Uniformity

Raw throughput uniformity of better than 75% for 1000 production units measured channel to channel
OEM Spectrometer Selection Guide

| General Spectrometer Specifications* | |
|---|---|
| 
 Spectral Coverage  | 
Tunable from 470 to 730 nm | 
| 
 Spectrometer Input  | 
 FC (standard), other input types available upon request  | 
| 
 Spectral Resolution  | 
2.7 nm with 75 µm slit; 5 nm for Ø 200 µm fiber (no slit) | 
| 
 Average Spectral Dispersion  | 
23.56 nm/mm; 0.33 nm/pixel (14 µm pixel CCD detector) | 
| 
 Crosstalk (channel to channel)  | 
< 0.05% (typical)+ < 0.1% (maximum)  | 
| 
 Focal Length  | 
70 mm | 
| 
 F/#  | 
~ F/2.3 | 
| 
 Stray Light Rejection (< 590 nm or > 680 nm)  | 
< 0.05 % (typical) | 
| 
 Wavelength Accuracy  | 
< 0.25 nm (average between 500 nm and 710 nm) | 
| 
 Software  | 
 LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration)  | 
| Back-illuminated linear CCD Detector | ||
|---|---|---|
| CCD Detector Model | Hamamatsu S11071-1104N | |
| CCD Sensor Format | 2048 x 22 (2 spectra per CCD of about 1000 pixels each) | |
| CCD Pixel Size | 14 x 14 µm | |
| CCD Height | 0.3 mm | |
| CCD QE at 20° C | QE = 63% at 250 nm; 76% at 650 nm; 55% at 850 nm | |
| Scan Rates | 2 MHz | |
| Full Well Capacity | > 225,000 e¯/pixel (minimum); 250,000 e¯/pixel (typical) | |
| Communication | USB 2 | |
| Scan Rates | 45 kHz and 500 kHz | |
| Readout Noise | 23 e¯ (typical) to 35 e¯ (maximum) | |
| Maximum Spectral Rate | 720 spectra/s with 0 ms exposure time | |
| Digitization | 16-bit ADC | |
| Dynamic Range (typical for single pixel)  | 
11,000:1 (typical) 7,300:1 (minimum)  | 
|
| Non-linearity (measured on each system)  | 
<0.8% (typical) 1.5% (maximum)  | 
|
| Dark Current at –45° C                                  Note: Pixel Size = 14 μm  | 
50 e¯/pixel/s (typical) | |
| Environmental Conditions | Operating temperature 15° C to 35° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C  | 
|
| Power requirements AC/DC power supply (provided)  | 
90-264 VAC, 47–63 Hz | |
* Specifications, form factor, and spectrometer cover subject to change without notice. 
No LabVIEW™ license is needed to run our acquisition software.
+ This value includes optical and electronic crosstalk